Unread postby ewanfowler » 08 Oct 2020, 16:02

Hi,

Can I please clarify what the update rate is for current being returned to the cell during dynamic clamp experiments with 1 model using a Power1401-3A? http://ced.co.uk/products/sigdcmp says 320 kHz for one HH model but I'm not sure if that's what's being sampled or the DAC output being updated. What would the sampling rate be in this case?

Cheers,
Ewan

User avatar
Tim Bergel
Site Admin
Posts: 2379
Joined: 19 Jun 2008, 14:40
Location: Cambridge, England
Contact:

Re: Dynamic clamp update rate

Unread postby Tim Bergel » 13 Oct 2020, 09:56

Hi Ewan,

The DAC updates during dynamic clamping are closely linked to the ADC sampling so as to minimise latency, what happens is that the dynamic clamp update routines are called immediately after a set of ADC samples is taken. So the ADC rate is the dynamic clamp update rate and you change both by setting the ADC rate.
Tim Bergel Cambridge Electronic Design

ewanfowler
Contributor
Posts: 47
Joined: 06 Sep 2016, 10:29
Software used: Signal
1401 type: Power1401
Contact:

Re: Dynamic clamp update rate

Unread postby ewanfowler » 13 Oct 2020, 12:21

Hi Tim,

Perfect thanks for the clarification!